国产亚洲VA色网,久久久久无码国产精品一区,亚洲阿v天堂无码在线,国产中文佬中文网无码av

ProductsYour current location:Home > Products
TR-2000 Mother/Fetus Monitor
Release time:2015-07-10 Author:admin Source:

 

 

Technical specification
FHR      Method: Ultrasound Doppler
              Measuring Range: 60-200bpm
              Precision: ≤±2bpm
              Ultrasound Frequency: 2.5 Mhz
              Ultrasound output: 5mW/cm2:
              Alarm Indication range: upper limit:66-200bpm
                                                            Lower limit: 65-199bpm
TOCO   Method: Outside Measurement
              Measuring Range: 0-100 unit
              Non-linearity Precision: ≤±10%
SPO2    SPO2 measuring range:0-100% Resolution:1%
              Range of pulse rate:20-250 bpm

NIBP     Method: Intelligent Oscillation
              Cuff inflating time: <30s
               (standard adult cuff : 0-250 mmHg)
               Deflating time: average<90sec
               Deflating time when measurement stops:<2sec
               (standard adult cuff 250-150mmmHg)
                Maximum Allowed Inflating Pressure: 300mmHg
               (40Kpa, if over this value, safety valve
         deflating automatically)
         Alarm indication range of FHR, TOCO, HR, NIBP
     SPO2 are settable. Sound and light alarm are available.
 

Coopyright 2015 www.jinzuoe.com Tianrong medical communication equipment Co., Ltd. All Rights Reserved
Company address: 21-1 Yangshan Road, Xuzhou Economic development zone, xuzhou city, jiangsu province, china, 24 hour Hotline :+86-13395229510 telephone :+86-516-85532818 E-mail : LB@tianrong.cn